Polychlorinated biphenyls (PCBs) are a class of persistent, bioaccumulative and toxic environmental pollutants that were banned from sale in the 1970s due to their detrimental impact on the environment.
Gas chromatography coupled to mass spectrometry (GC–MS) is considered an extremely reliable tool for PCB analysis but typically relies on helium as a carrier gas. This leaves laboratories reliant on an increasingly expensive and difficult-to-source gas.
This application note demonstrates how hydrogen carrier gas can be used in place of helium for GC-MS/MS analysis of PCBs, giving laboratories greater flexibility and ensuring sustained, cost-effective operations.
Download this application note to discover:
- How hydrogen can replace helium without compromising analytical performance
- A unique solution for minimizing carrier gas consumption
- How to reduce costs and enable longer operational times on a single gas tank
Goal
The aim of this application note is to compare the performance of the Thermo Scientific™
TSQ™ 9610 triple quadrupole mass spectrometer coupled to the Thermo Scientific™
TRACE™ 1610 GC with hydrogen versus helium as the carrier gas for the determination
of polychlorinated biphenyls. For guidance on the analytical performance of the
proposed method, acceptance criteria as per U.S. EPA Method 1668 were considered.
Introduction
Polychlorinated biphenyls (PCBs) are a group of industrial organic chemicals
consisting of carbon, hydrogen, and chlorine atoms. Due to their non-flammability,
chemical stability, high boiling point, and electrical insulating properties, PCBs
were used in hundreds of industrial and commercial applications including
electrical, hydraulic equipment, plasticizers, plastics, rubber products, and dyes.
The production of these compounds has been banned in the United States
since 19771
because of their persistence in the environment and their tendency to
enter the food chain and bioaccumulate in living organisms due to their lipophilicity.
Environmental analysis of polychlorinated biphenyls (PCBs)
at reduced running costs using hydrogen as carrier gas
Application note | 002549
Keywords
Polychlorinated biphenyls, PCBs,
HeSaver-H2
Safer, GC-MS/MS,
H2
carrier gas, U.S. EPA 1668
Environmental
Authors
Łukasz Rajski, Daniel Kutscher
Thermo Fisher Scientific, Bremen, Germany
There are currently 209 known PCBs congeners that can
be divided into two groups according to their structural and
toxicological characteristics:
• Non-dioxin-like PCBs (non-DL-PCBs), which represent the
majority of the PCB congeners, are characterized by a lower
degree of toxicity.
• Dioxin-like PCBs (DL-PCBs) include the 12 most toxic
congeners (non‐ortho PCBs 77, 81, 126, 169 and mono‐ortho
PCBs 105, 114, 118, 123, 156, 157, 167, 189), which have
structures and toxicities similar to dioxins.
DL-PCBs are classified as persistent organic pollutants (POPs),
and they have been regulated under the Stockholm Convention
for POPs since 2001.2
Following the Clean Water Act (CWA) in
1972, the United States Environmental Protection Agency
(U.S. EPA) developed an analytical method, U.S. EPA
Method 1668 and following revisions3
that can be applied
for the determination of PCBs in wastewater, surface waters,
soil, sediments, biosolids, and tissue matrices using gas
chromatography coupled to high-resolution gas chromatography/
high-resolution mass spectrometry (HRGC/HRMS).3
However,
recent advances in gas chromatography-triple quadrupole mass
spectrometry allow for high sensitivity and selectivity, leading
to the consideration of GC-MS/MS as a reliable tool for PCBs
analysis.
This application note reproduces the experiments described in
a previously published application note,4
where polychlorinated
biphenyls were analyzed using helium carrier gas. Here, helium
was replaced by hydrogen. Additionally, the Thermo Scientific™
HeSaver-H2
Safer™ carrier gas saving technology was applied.
HeSaver-H2
Safer technology offers a unique solution for
laboratories to minimize carrier gas consumption during both
standby and operational modes. When helium is used as carrier
gas, the consumption can be drastically reduced without any
changes to the analytical method or deterioration of performance.
However, the Split-Splitless (SSL) injector modified to work in
the HeSaver-H2
Safer mode can be used also in conjunction with
hydrogen as a carrier gas, where the limited and fixed carrier gas
flow allows for safe usage without the need to install additional
sensors. At the same time, the hydrogen gas consumption is
equally reduced and will lead to further cost savings, allowing
laboratories to run their instrumentation longer on a single gas
tank or using gas generators.
Experimental
The analytical method used for this study is described in a
previous application note,4
however, minor modifications to
the method were made and are summarized in Table 1. The
use of hydrogen as carrier gas often leads to shorter retention
times, if the analytical conditions, such as column, carrier gas
flow rate, and oven program, remain unchanged. The retention
times obtained using hydrogen as carrier gas can be found in
Appendix 1.
Table 1. GC-MS parameters modified in comparison to AN000561
SSL parameters
Inlet module and mode SSL, HeSaver-H2
Safer, Splitless
Split flow (mL/min) 50 (Nitrogen)
Carrier gas, flow (mL/min) H2
; 1.2
Pressurizing gas Nitrogen
Oven
Column Thermo Scientific™ TRACE™
TR-PCB 8 MS; 50 m, 0.25 mm, 0.25 μm
(P/N 26AJ148P)
TSQ 9610 mass spectrometer parameters
Transfer line temperature (°C) 330
Ion source type and
temperature (°C) AEI, 350
Emission current (μA) 10
Results and discussion
Separation
Maintaining chromatographic resolution is critical when analyzing
PCBs. In the analyzed set of PCBs, the pentachlorobiphenyls,
commonly referred to as PCB-123 and PCB-118, were monitored.
These two compounds have similar retention times and identical
transitions. Therefore, a good chromatographic method is
necessary to avoid interference. With the TRACE TR-PCB 8
MS column, baseline separation of these two congeners was
achieved (Figure 1) with no modification or amends to the
chromatographic method.
Instrumental detection limits
One of the main concerns when changing the carrier gas
from helium to hydrogen is a potential sensitivity drop. The
instrumental detection limits (IDLs) were determined for the
individual congeners by diluting the CS0.2 standard four times,
so that a final concentration of 0.05 ng/mL was achieved. This
solution was repeatedly injected (n=10). IDLs were calculated
considering the one-tailed Student’s t-test values for the
corresponding n-1 degrees of freedom at 99% confidence, the
injected on-column (OC) concentration, and the absolute peak
area %RSD (<15%) for each analyte. Figure 2 shows that the
IDLs obtained with hydrogen were nearly identical to those with
helium. All compounds had an IDL under 20 fg except PCB-209.
2
Figure 1. Quantifying peaks of all analyzed PCBs (A) and separation of the critical pair of PCB congeners 123/118 (B)
0
5
10
15
20
25
30
PCB-1
PCB-3
PCB-4
PCB-19
PCB-15
PCB-54
PCB-104
PCB-37
PCB-155
PCB-77
PCB-81
PCB-188
PCB-123
PCB-118
PCB-114
PCB-105
PCB-126
PCB-202
PCB-167
PCB-156
PCB-157
PCB-169
PCB-208
PCB-189
PCB-205
PCB-206
PCB-209
fg OC
Instrumental detection limits
H2 He
Figure 2. Instrumental detection limits expressed as the mass of PCB
injected on column
8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
0.0e0
2.5e8
5.0e8
7.5e8
1.0e9
1.3e9
1.5e9
1.8e9
2.0e9
12.600 12.625 12.650 12.675 12.700 12.725 12.750 12.775 12.800 12.825 12.850 12.872
0.0e0
5.0e7
1.0e8
1.5e8
2.0e8
2.5e8
3.0e8
3.5e8
4.0e8
4.5e8
5.0e8
5.5e8
A)
B)
PCB-123
PCB-118
Counts
Minutes
Minutes
Counts
3
ratio of 10 injections at 0.05 ppb. The reference values for the
data in Figure 3B were the average ion ratios from the calibration
curve.
Linearity
All investigated PCBs showed excellent linearity. All analytes
achieved a coefficient of determination equal to or higher than
0.9998 (Figure 4A), and the relative standard deviation of the
average response factor did not exceed 3% (Figure 4B).
Ion ratio
The correct ion ratio is very important for the unambiguous
identification of the analytes. However, at low concentration
levels it can deviate from the expected value and subsequently
generate a false negative result. The reproducibility of the
ion ratio was thus investigated across a wide concentration
range. Figure 3A presents the ion ratio values of PCB-114 as an
example, whereas Figure 3B depicts a deviation of the average
Figure 3. Ion ratio of PCB 114 at various concentration levels (A) and ion ratio deviation of all analytes at 0.05 ppb (B)
Figure 4. Evaluation of the linearity in the range 0.1–2000 ppb: coefficient of determination (A) and relative standard deviation of the
average response factor (B)
0%
20%
40%
60%
80%
100%
120%
140%
PCB 114
Ion ratio Average ratio + 20% Average ratio - 20%
0%
2%
4%
6%
8%
10%
12%
PCB-1
PCB-3
PCB-4
PCB-19
PCB-15
PCB-54
PCB-104
PCB-37
PCB-155
PCB-77
PCB-81
PCB-188
PCB-123
PCB-118
PCB-114
PCB-105
PCB-126
PCB-202
PCB-167
PCB-156
PCB-157
PCB-169
PCB-208
PCB-189
PCB-205
PCB-206
PCB-209
Ion ratio deviation
H2 He
A)
0.1ppb 0.2 ppb 1 ppb 5 ppb 50 ppb 400 ppb 2000 ppb
B)
0.9900
0.9920
0.9940
0.9960
0.9980
1.0000
PCB-1
PCB-3
PCB-4
PCB-19
PCB-15
PCB-54
PCB-104
PCB-37
PCB-155
PCB-77
PCB-81
PCB-188
PCB-123
PCB-118
PCB-114
PCB-105
PCB-126
PCB-202
PCB-167
PCB-156
PCB-157
PCB-169
PCB-208
PCB-189
PCB-205
PCB-206
PCB-209
Coecient of determination R2
H2 He
0.0%
2.0%
4.0%
6.0%
8.0%
10.0%
12.0%
14.0%
16.0%
18.0%
20.0%
PCB-1
PCB-3
PCB-4
PCB-19
PCB-15
PCB-54
PCB-104
PCB-37
PCB-155
PCB-77
PCB-81
PCB-188
PCB-123
PCB-118
PCB-114
PCB-105
PCB-126
PCB-202
PCB-167
PCB-156
PCB-157
PCB-169
PCB-208
PCB-189
PCB-205
PCB-206
PCB-209
RSD AvCF
H2 He
A) B)
4
General Laboratory Equipment – Not For Diagnostic Procedures. © 2023 Thermo Fisher Scientific Inc. All rights
reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. This information is
presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products
in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not
all products are available in all countries. Please consult your local sales representative for details. AN002549-EN 1023S
Learn more at thermofisher.com/helium-saver
Summary
The HeSaver-H2
Safer inlet allows laboratories full flexibility to
select the carrier gas of choice between helium and hydrogen,
while achieving a significant reduction of the consumption. With a
simple upgrade to the standard Thermo Scientific™ iConnect SSL
injector, the HeSaver-H2
Safer inlet allows for the use of an inert
and inexpensive gas for inlet pressurization, whereas the main
carrier gas flow is limited to the column flow only.
• The use of hydrogen as carrier gas has been demonstrated as
an alternative to helium for the analysis of PCBs, reducing the
overall cost of running the laboratory.
• The use of the HeSaver-H2
Safer inlet reduces the amount of
hydrogen going into the gas chromatography system such
that safe operation is assured anytime, and no hydrogen
sensor is needed inside the GC oven.
• Hydrogen did not affect the linearity range; all compounds
showed a linear response from 0.1 to 2000 ppb, with avarage
R2
>0.999.
• The TSQ 9610 triple quadrupole mass spectrometer provided
excellent IDLs with hydrogen carrier gas, comparable to those
obtained with helium.
• The variation of ion ratios was less than 12% for all PCBs at
0.05 ppb.
References
1. United States Environmental Protection Agency, U.S. EPA, Learn
about Polychlorinated Biphenyls (PCBS), https://www.epa.gov/pcbs/
learn-about-polychlorinated-biphenyls-pcbs
2. Stockholm Convention. http://chm.pops.int/Implementation/IndustrialPOPs/PCBs/
Overview/tabid/273/Default.aspx
3. United States Environmental Protection Agency, U.S. EPA, Method 1668C -
Chlorinated Biphenyl Congeners in Water, Soil, Sediment, Biosolids, and Tissue by
HRGC/HRMS, April 2010, https://www.epa.gov/sites/default/files/2015-09/documents/
method_1668c_2010.pdf
4. Thermo Fisher Scientific, Application Note 000561: Reproducible trace analysis of
PCBs in environmental matrices using triple quadrupole GC-MS/MS. https://assets.
thermofisher.com/TFS-Assets/CMD/Application-Notes/an-000561-gc-ms-water-soilpcbs-an000561-en.pdf
Appendix 1. List of PCBs retention times
Compound
Retention time
[min] Compound
Retention time
[min] Compound
Retention time
[min]
PCB-1 7.61 PCB-77 12.04 PCB-167 15.02
PCB-3 8.04 PCB-81 12.34 PCB-156 15.79
PCB-4 8.16 PCB-188 12.53 PCB-157 15.99
PCB-19 8.72 PCB-123 12.69 PCB-169 17.24
PCB-15 9.11 PCB-118 12.79 PCB-208 17.64
PCB-54 9.27 PCB-114 13.03 PCB-189 18.50
PCB-104 9.99 PCB-105 13.56 PCB-205 19.79
PCB-37 10.43 PCB-126 14.60 PCB-206 20.44
PCB-155 10.84 PCB-202 14.70 PCB-209 20.93