Improve Your Environmental Analysis With a Single GC-MS/MS Workflow
App Note / Case Study
Published: July 2, 2024
Credit: iStock
In the evolving landscape of environmental monitoring, advanced analytical techniques are essential for accurately detecting different types of contaminants.
To do this, traditional methods often require multiple instruments, resulting in long turnaround times, increased costs and inefficiencies.
This application showcases a simple GC-MS/MS workflow that can be used to analyze different environmental contaminants and can help environmental laboratories to maximize efficiency, productivity and revenue.
Download this application note to explore:
- An innovative GC-MS/MS approach that enhances productivity, reduces costs and simplifies operations
- Detailed experimental setups to analysze polybrominated diphenylethers (PBDEs) pesticides, and micropollutants
- How to meet regulatory requirements for different compound classes
Goal
This application note will highlight a single instrument configuration, based on triple
quadrupole GC-MS, to perform the analysis of various compound classes commonly
monitored in the environment. This will allow laboratories to standardize on a single
hardware configuration and hence greatly facilitate laboratory capacity, instrument- and
consumables management, as well as streamlining staff training.
Introduction
Analytical testing laboratories play a pivotal role in monitoring common contaminants
in different matrices. While data quality and compliance with established regulated
methods is key in ensuring a laboratory’s reputation, productivity and fast sample
turnover times are equally important to ensure profitable operation.
In the field of environmental monitoring, a large variety of compound classes may be
required for analysis, including both common and emerging contaminants. For volatile
and semi-volatile substances, gas chromatography is the method of choice for separating
pesticides, polyaromatic hydrocarbons (PAHs), flame retardants such as polybrominated
diphenyl ethers (PBDEs), and others. In many cases, due to previously existing
methodologies, samples need to undergo several analyses on different instrument
configurations, with either specific consumables, including the analytical column or a
dedicated detector. This leads to a variety of protocols and potential instrument setups
required in a laboratory, some of which must be run in a dual-column configuration or a
confirmation on a separate instrument using a second injection is required.
Improved laboratory productivity with a single GC-MS/MS
configuration for multipurpose environmental analysis
Application note | 002866
Authors
Delphine Thomas, Bénédicte Gauriat
Thermo Fisher Scientific Customer Solution
Center, Les Ulis, France
Keywords
Environmental analysis, pesticides,
PBDEs, PCBs, micropollutants,
GC-MS/MS, method consolidation
Environmental
While mass spectrometric detection provides a much higher
selectivity than analog detectors, only triple quadrupole systems
provide the selectivity needed to effectively leverage selected
rection monitoring (SRM) as a tool for method standardization and
consolidation. The power of triple quadrupole mass spectrometry
can be easily used to modernize existing workflows. Laboratories
can reduce their need for instrument-specific consumables and
components on-hand to reduce costs and complexity with a
single instrument configuration. This application demonstrates the
use of a single hardware configuration for the analysis of different
environmental contaminants.
Experimental
For all measurements, a Thermo Scientific™ TRACE™ 1610 GC
coupled to Thermo Scientific™ TSQ™ 9610 GC-MS/MS was used.
The system was equipped with a Thermo Scientific™ iConnect™
Thermospray SSL Injector Module (TSI).
The analytical column used for all substance classes was a
Thermo Scientific™ TraceGOLD™ TG-Contaminants GC column,
which allows for fast analysis of multi-class contaminants
including pesticides, polychlorinated biphenyls (PCBs), PAHs,
and PBDEs. These highly inert capillary columns combine
thermal stability with a unique selectivity to ensure excellent
chromatographic peak shapes and sensitivity and allow
chromatography of semi-volatile compounds together with lateeluting compounds. Table 1 gives a high-level summary of the
parameters for each of the three methods showcased.
Single GC-MS/MS configuration
• Thermo Scientific™ TriPlus™ RSH SMART autosampler
(P/N 1R77010-2003)
• TRACE GC 1610 (P/N MI-148000-0007)
• Thermospray injector TSI (P/N 19070016)
• TSQ 9610 AEI source (P/N TSQ9610-NV-AEI)
• TraceGOLD TG-Contaminants GC column
15 m × 0.25 mm × 0.1 µm (P/N 26056-0350)
• Thermo Scientific™ LinerGOLD™ splitless precision liner,
quartz wool (P/N 453T2999-UI)
Figure 1 gives detailed information on the configuration used for
the analysis.
Table 1. Instrument parameters
For further information, contact your local commercial representative
Parameter PBDE Pesticides Micropollutants
Injection mode Splitless Split with surge Split
Injection volume (µL) 2 µL 3 µL 2 µL
GC run time (min) 16 minutes 20 minutes 20 minutes
Number of compounds monitored 9 350 220
Calibration range in solvents (µg/L) 1 to 500 2 to 250 2 to 250
Compound classes covered • Flame
retardants
Acquisition mode SRM
•Carbamates
•Triazines
•Azoles
•Ureas
•Organochlorides
•Organophosphorous
•Organonitrogens
•Pyrethroids
•PCBs
•Phenolic derivatives
•PAHs
•Aniline derivatives
•Benzene derivatives
•Morpholines
•Anilides
•Uracils
•Amides
•Strobilurins
TSQ 9610 mass spectrometer with the Advanced Electron Ionization
(AEI) source
TRACE 1610 GC with Dedicated TraceGOLD TG-Contaminants
GC column
Intelligent software allows
AutoSRM and dwell time
optimization to ensure peak
analytical performance.
Modular design helps
reduce downtime and
simplifies maintenance
requirements.
Dedicated TraceGOLD
TG-Contaminants GC column
allows multiple compounds
to be analyzed on a single
column chemistry.
Thermospray SSL
Injector Module enables
larger sample volume injections
due to its large spray chamber and
helps to improve the sensitivity of
labile compounds.
Using NeverVent
technology, maintenance
can be performed without
breaking vacuum.
AEI source sensitivity
allows regulatory limits
to be met with ease.
XLXR™ detector aids in
method consolidation with
its wide dynamic range and
less maintenance due to its
extended lifetime.
Figure 1. Configuration utilized for analysis the TRACE 1610 GC with the TSQ 9610 NV-AEI GC-MS/MS
2
Results and discussion
Analysis of polybrominated diphenylethers (PBDE)
PBDEs are used as additive flame retardants in different materials
and can leach into the environment where they persist and
bioaccumulate.1
PBDEs are a class of brominated hydrocarbons
with a basic structure containing two phenyl rings linked by
an oxygen atom. Within the European Standard 16694:2025,
applicable for water analysis, the determination of six selected
PBDEs is required, which include congeners BDE-28, -47, -99,
-100, -153, and -154.
The proposed setup has been tested to comply with the
requirements, using the conditions summarized in Table 1.
Figure 2. SRM plot for the separation of PBDEs at a concentration of 1 ppm, in line with the requirements of European Standard 16694:2015
6.580 6.625 6.750 6.875 6.980
0.0e0
1.3e5
2.5e5
3.8e5
5.0e5
6.0e5
min
counts
BDE028
9.090 9.250 9.375 9.490
0.0e0
2.0e4
4.0e4
6.0e4
8.0e4
1.0e5
min
counts
BDE183
2.5e5
counts
BDE099
6.580 6.625 6.750 6.875 6.980
0.0e0
1.3e5
2.5e5
3.8e5
5.0e5
6.0e5
min
counts
BDE028
9.090 9.250 9.375 9.490
0.0e0
2.0e4
4.0e4
6.0e4
8.0e4
1.0e5
min
counts
BDE183
7.940 8.000 8.125 8.250 8.340
0.0e0
5.0e4
1.0e5
1.5e5
2.0e5
2.5e5
min
counts
BDE099
6.580 6.625 6.750 6.875 6.980
0.0e0
1.3e5
2.5e5
3.8e5
5.0e5
6.0e5
min
counts
BDE028
9.090 9.250 9.375 9.490
0.0e0
2.0e4
4.0e4
6.0e4
8.0e4
1.0e5
min
counts
BDE183
7.940 8.000 8.125 8.250 8.340
0.0e0
5.0e4
1.0e5
1.5e5
2.0e5
2.5e5
min
counts
Figure 3. Overlaid SRM traces at 1 µg/L (quantifier and qualifier) BDE099
7.5 8.5 9.5 10.5 11.5
Time [min]
7.0 8.0 9.0 10.0 11.0 12.0 12.5
6.2e8
Counts 5.0e8
4.0e8
3.0e8
2.0e8
1.0e8
0.0e0
1.1e9
1.0e9
9.0e8
8.0e8
7.0e8
In addition to the required congeners, BDEs 183, 207, and 209
also have been monitored; they can be separated and quantified
at required levels, as is demonstrated by the selected reaction
monitoring plot shown in Figure 2. Figure 3 shows a more detailed
SRM plot for BDE-28, BDE-99, and BDE-183. These plots show
that four SRM transitions are clearly visible at this concentration,
which aids in confirming the presence of the compounds in an
environmental sample.
To demonstrate the performance, the chromatograms shown in
Figure 3 present three examples of PBDE compounds at 1 µg/L
in solvent using SRM analysis.
3
and physical properties but can be grouped into multiple subclasses, i.e., organophosphorus compounds, pyrethroids,
organochlorides, carbamates, and azoles.
Using the proposed methodology, described in more detail
elsewhere,2,3 up to 350 individual pesticides can be analyzed,
covering a wide range of compound classes. A full overview of
the entire separation is given in Figure 4.
Figure 5 shows examples of SRM chromatograms for five
pesticide compounds (class of organophosphorus, pyrethroid,
organochloride, carbamate, and azole) at a concentration of
3 µg/L.
Figure 4. Two SRM plots of two mixtures of pesticides covering 350 compounds at a concentration of 10 ppb, showing
good separation within 20 minutes
Analysis of pesticides
Even though pesticides are commonly analyzed in food samples,
they also need to be determined in water samples to ensure that
they do not contaminate waterways and in turn have a negative
effect on the ecosystem. One example of a regulation to monitor
pesticides in water is the European Groundwater Directive
(2006/118/EC).
Approximately 1,000 active substances are currently used in
pesticide formulations. In addition, metabolites, degradation
products, and ‘‘old’’ (persistent) pesticides must be considered
by pesticide residue analysts. Pesticides have different chemical
Figure 5. Selected pesticides chromatograms at 3 µg/L
579 11 13 16 17
Time [min]
4 6 8 10 12 14 15
5 7 9 11 13 16 17
Time [min]
4 6 8 10 12 14 15
1.2e7
Counts
1.0e7
8.0e6
6.0e6
4.0e6
2.0e6
0.0e0
Pesticides mix 2
Pesticides mix 1
1.2e7
Counts
1.0e7
8.0e6
6.0e6
4.0e6
2.0e6
0.0e0
579 11 13 16 17
Time [min]
4 6 8 10 12 14 15
5 7 9 11 13 16 17
Time [min]
4 6 8 10 12 14 15
1.2e7
Counts
1.0e7
8.0e6
6.0e6
4.0e6
2.0e6
0.0e0
Pesticides mix 2
Pesticides mix 1
1.2e7
Counts
1.0e7
8.0e6
6.0e6
4.0e6
2.0e6
0.0e0
4
An example of an SRM chromatogram for the full method
including all the classes under evaluation is given in Figure 6.
In Figure 7, chromatograms summarize five representive
examples across all compound classes included in the method,
namely derivatives of phenolics, benzenes, anilines, PCBs, and
PAHs. Each compound was present in the mix at a concentration
of 5 µg/L. 2-chlorophenol, nitrobenzene, pentachloroaniline,
1,5-dichloronaphthalene, and PCB209 were selected as
representative compounds of the compound classes.
Analysis of micropollutants
The term micropolluntants describes different, mostly organic,
chemicals that may enter the environment from anthropogenic
processes. A main source for micropollutants is wastewater from
all sources, including industrial plants, agricultural processes, and
private households. Even though micropollutants are present in
low concentrations, long-term exposure may pose considerable
risks to aquatic organisms and human health. Micropollutants can
be separated into a variety of compound classes, out of which a
great number are amenable to gas chromatographic analysis.
Figure 6. SRM plot of the micropollutants full method at a concentration of 50 ppb
Figure 7. Overlaid SRM traces at 5 µg/L (quantifier and qualifier)
3.074 3.200 3.400 3.474
-1.0e4
0.0e0
2.0e4
4.0e4
6.0e4
8.0e4
1.0e5
min
counts
2-CP
4.021 4.200 4.421
-5.0e4
0.0e0
1.0e5
2.0e5
3.0e5
4.0e5
4.5e5
min
counts
Nitrobenzene
9.061 9.200 9.400 9.461
-2.0e5
0.0e0
2.5e5
5.0e5
7.5e5
1.0e6
1.3e6
1.4e6
min
counts
1,5-Dichloro-Naphtalene
10.601 10.800 11.001
-2.0e4
0.0e0
5.0e4
1.0e5
1.5e5
1.8e5
min
counts
Pentachloroaniline
13.785 14.000 14.185
-5.0e3
0.0e0
1.0e4
2.0e4
3.0e4
4.0e4
4.5e4
min
counts
PCB 209
2468 10 12 14 16
2.4e7
Time [min]
Counts
2.0e7
1.6e7
1.2e7
8.0e6
4.0e6
0.0e0
5
Figure 8. Peak area %RSD obtained for the QC checks at 5 μg/L (n=7), analyzed before and after
the repeated injection of matrix samples (n=80)
6%
6%
8%
9%
5%
0%
1%
2%
3%
4%
5%
6%
7%
8%
9%
10%
0
5000
10000
15000
20000
25000
2-Chlorophenol
Nitrobenzene
Pentachloroaniline
1,5-Dichloro-Naphthalene
PCB209
RSD
%
Area ratio
Target compounds
QC5µg/L-1 AREA QC5µg/L-2 AREAO QC5µg/L-3 AREA QC5µg/L-4 AREA
QC5µg/L-5 AREA QC5µg/L-6 AREA QC5µg/L-7 AREA
Robustness and repeatability of the area at 5 µg/L
To illustrate the robustness and repeatability of the method, the
average peak area determined for all compounds covered in the
method was evaluated. This was accomplished using QC check
standards at 5 µg/L, injected before and after the analysis of
80 water samples. No maintenance or tuning was performed on
either the inlet system or the mass spectrometer. The results are
displayed in Figure 8.
Conclusion
The utilization of a single configuration GC-MS/MS for multiple
environmental methods has been demonstrated in this application
note. As well as being able to meet the regulatory requirements
for different compound classes, there are advantages to using a
single configuration for multiple analysis:
• Reduction in operation cost, including reduced training needs
and centralization of consumables
• Ability to react and be ready to run the required method
without lengthy set up or changing instrument analytical
column and hardware
• Increase in instrument capacity as all systems can run all
methods
The Thermo Scientific GC-MS/MS system also increases
productivity:
• The TRACE 1610 GC has a unique module design allowing
injectors and detectors to be exchanged by the user and
maintenance to be performed rapidly offline. The TSI injector
allows a wider range of compound classes with different
volatilities to be covered in a single injection. The easy-to-use
touch screen provides real-time monitoring of the system and
video tutorials for quick adoption.
• The TSQ 9610 GC-MS/MS is equipped with Thermo
Scientific™ NeverVent™ technology, which allows the GC
column to be exchanged, the ionization source to be
removed, and filaments to be removed without the need for
breaking vacuum. This allows routine maintenance to be
performed in minutes and increases system uptime drastically.
In summary, this single GC-MS/MS configuration can greatly
simplify the workflows of environmental laboratories and increase
efficiency and productivity, reduce costs, and maximize revenue.
6
General Laboratory Equipment – Not For Diagnostic Procedures. © 2024 Thermo Fisher Scientific Inc. All rights
reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. This information is
presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products
in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not
all products are available in all countries. Please consult your local sales representative for details. AN002866-EN 0424S
Learn more at thermofisher.com/tsq9610
References
1. Hites, R.A. Polybrominated diphenyl ethers in the environment and in people: a
meta-analysis of concentrations. Environ. Sci. Technology 2004, 38, 945.
2. Thermo Fisher Scientific Application Brief AB00019: Improving the routine testing
of difficult-to-analyze pesticide residues by using the Thermo Scientific TraceGOLD
TG-Contaminants column with GC-MS/MS technology. https://assets.thermofisher.
com/TFS-Assets/CMD/Application-Notes/ab-000019-gcms-div-ctl-tg-contaminantsab000019-en.pdf
3. Thermo Fisher Scientific Application Note AN00780: A robust and sensitive method
for the determination of pesticides in surface and ground water by triple quadrupole
GC-MS. https://assets.thermofisher.com/TFS-Assets/CMD/Application-Notes/
an-000780-gc-ms-tsq-9610-pesticides-groundwater-an000780-na-en-1.pdf
Brought to you by
Download this App Note for FREE Below!
Information you provide will be shared with the sponsors for this content.
Technology Networks or its sponsors may contact you to offer you content or products based on your interest in this topic. You may opt-out at any time.
Experiencing issues viewing the form? Click here to access an alternate version